348
Trigger mode
Open the trigger configuration menu and select the appropriate trigger type. When the I2C bus trigger is
selected, click the trigger type and relationship on the screen, as shown in Figure 12-23:
Figure 12-23 I2C Trigger Mode Configuration Menu
Trigger mode menu description:
a) Start condition — trigger when SCL is high and SDA has a falling edge (includes restart).
b) Stop condition — trigger when SCL is high and SDA has a rising edge.
c) Ack Loss — Trigger when the bus Ack bit is high.