Maintenance
63
Table 7. Error Messages (cont.)
Error Message Explanation Potential Root Cause
Filesystem fatal error
Failed to load filesystem from flash
memory when startup.
Data flash memory failure, the main PCA
needs service.
Cannot save screen Failed to save screen into filesystem.
Not enough space, delete some files and
try again.
Calibration error when
(name of step)
Calibration failed at current step.
Make sure connection and reference are
correct and try again.
HART command failed (command
dependent explanation)
Error returned from HART device of
previous command.
Device-dependent error, please refer to
user manual of HART transmitter.
Lost connection to PCM The link to the PCM is broken.
Cycle power, if error persists, the unit
needs service.
Lost connection to EMM The link to EMM is broken.
PCM temperature too high
The temperature of PCM reaches its high
limit.
In measure mode, let the PCM cool
down and try the task again.
1.888.610.7664 sales@GlobalTestSupply.com