Automatic Pressure Calibrator
Tasks Menu
29
Pressure Switch (SW)
Use this feature to calibrate a pressure switch. See
Figure 12 for connections:
1. From page 1 of the Tasks menu, push the arrows to
highlight Pressure Switch (SW).
2. Push E or R.
3. From this screen, use the arrows, numeric keypad,
and E to change the values of these parameters:
Setpoint
Tolerance
Min Deadband
Max Deadband
Max Pressure
Switch Type
Note
Make sure to push E after EACH value
change.
You can save these settings as a customized task.
Push B (Save as Customized). Recall this
customized task from the Tasks menu.
4. Once you enter the parameters, push D
(Continue).
5. Push D (Auto Test) or B (Manual Test)
to do the As Found switch test. The test runs.
After the test runs, the results show on the display.
To save the test results:
1. Push D (Done).
2. Push E to edit the Tag information.
3. Push D (Done).
4. Move the cursor to the S/N field (Serial Number).
5. Push E to edit the S/N information.
6. Push D (Done).
7. Move the cursor to the User ID field.
8. Push A (User List) to select a user.
9. Push D (Done).
10. Adjust the switch and push B(Step Up)
and C(Step Down) to verify.
11. Push A (As Left) to do the As Left test.
1.888.610.7664 sales@GlobalTestSupply.com