729/729FC
Users Manual
54
Log
Users can record a series of pressure or mA
measurements for later upload to a host computer. The
Product records a maximum of 8000 readings, depending
on the reading rate, duration, and how much memory is in
use for other functions such as tasks or stored results.
Record the reading rate and duration in minutes.
To log data:
1. From the main screen, push B (Log). Adjust
the Reading Rate and Duration with the arrow
buttons and then push E to select parameters
from the Log screen.
2. Choose the Log Source from any of the available
measured values with A, B, or C.
The log source can be internal pressure, external
pressure, mA, V dc, HART PV, or temperature.
3. For Reading Rate: push E to adjust the reading
rate from 1 /m to 60 /m.
4. For Duration: use the numeric keypad to record the
duration in minutes, followed by E. The maximum
depends on the reading rate and how much memory
is available to log data.
5. Push D (Start) to begin. The Product logs the
measured values of the selected source for the pre-
selected duration. Push A (Abort) to stop
logging.
Results show after the Product logs the data. Push
A (Abort) or D (Done). If you push D
(Done), enter Tag, S/N, and a User ID. The Product
saves the results and are recalled under the
Setup>Manage Test Results menu. Upload the logged
Product data to DPCTrack2 for review in a graphical
report.
Communication with a PC
Download procedures from a PC to the Product and
upload test results to a PC from the Product. A PC,
Microsoft Windows, USB cable (supplied), and Fluke
DPCTrack2
application software, or a qualified Fluke
partner’s software are required. See the DPCTrack2
Users Manual for further instructions. See Figure 22 for