729/729FC
Users Manual
24
Manage Custom Tasks
Use this submenu to delete custom tasks (see Tasks
Menu). To delete a single task, use the arrow to select a
custom task and push A (Delete). To delete all of
the tasks, push D (Delete All).
Maintenance Menu
Maintenance has these sections:
• Drain Water (condensation)
• Keypad Test
• Pressure Sensor Characterize (see Calibration
Manual)
• Supply Sensor Self Calibration (see Calibration
Manual)
• Exhaust
Drain Water (Condensation)
To drain condensation from the Product, see Figure 10:
1. Connect the hose to collect drained water.
2. Push D (Continue).
3. The display shows Creating pressure, please
standby. The pump works to establish a pressure.
4. To let the water out, use a small-tipped tool, such as
a small Philips screwdriver, to push the DRAIN
button on the side of the Product.
5. Push D to repeat the drain operation.
6. When finished with water drain, push in on the fitting
to release the hose.