6 SHA800A QuickGuide
Power and Grounding Requirements
The instrument operates with a single-phase, 100 to 240 Vrms (+/-10%) AC power
at 50/60 Hz (+/-5%), or single-phase 100 to 120 Vrms (+/-10%) AC power at 400 Hz
(+/-5%).
No manual voltage selection is required because the instrument automatically
adapts to line voltage.
Depending on the type and number of options and accessories (probes, PC port
plug-in, etc.), the instrument can consume up to 193 W of power.
Note: The instrument automatically adapts to the AC line input within the following
ranges:
Voltage Range:
90 - 132 Vrms
Frequency Range:
380 - 420 Hz
The instrument includes a grounded cord set containing a molded three-terminal
polarized plug and a standard IEC320 (Type C13) connector for making line
voltage and safety ground connection. The AC inlet ground terminal is connected
directly to the frame of the instrument. For adequate protection against electrical
shock hazards, the power cord plug must be inserted into a mating AC outlet
containing a safety ground contact. Use only the power cord specified for this
instrument and certified for the country of use.
Warning: Electrical Shock Hazard!
Any interruption of the protective conductor inside or outside of the
analyzer, or disconnection of the safety ground terminal creates a
hazardous situation. Intentional interruption is prohibited.
The position of the instrument should allow easy access to the socket. To make the
instrument completely power off, unplug the instrument power cord from the AC
socket.