Chapter 1 Quick Start RIGOL
MSO8000 User's Guide 1-9
Function I nspection
1. Press Default on the front panel, then a prompt message "Restore default
settings?" is displayed. Press OK or tap OK to restore the instrument to its
factory default settings.
2. Connect the ground alligator clip of the probe to the "Ground Terminal" as
shown in Figure 1-10 below.
3. Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe, as shown in Figure 1-10.
Figure 1-10 To Use the Compensation Signal
4. Set the probe attenuation to 10X, and then press AUTO.
5. Observe the waveform on the display. I n normal condition, the square waveform
as shown in Figure 1-11 should be displayed.
Figure 1-11 Square Waveform Signal
6. Use the same method to test the other channels. I f the square waveforms
actually shown do not match that in the f igure above, please perform "Probe
Compensation" introduced in the next section.
WARNI NG
To avoid electric shock when using the probe, please make sure that the
insulated wire of the probe is in good
condition. Do not touch the
metallic part of the probe when the probe is connected to high voltage
source.
Compensation Signal Output Terminal
Ground Terminal