WWW.SIGLENT.COM 90
Command format: [:SOURce]:OUTPut:LOCAL:ENable? <channel>
Description: Get the external output enable status of the selected channel
E.g: OUTPut:LOCAL:ENable? CH1
Response: 1\n
Set internal resistance
Command format: [:SOURce]:INIT:RESIstance <channel>,{<value> | MINimum | MAXimum |DEFault}
Description: Set the internal resistance of the selected channel
E.g: INIT:RESIstance CH1,2
Description: Set the internal resistance of CH1 channel to 2Ω
Command format: [:SOURce]:INIT:RESIstance? <channel>
Description: Get the internal resistance value of the selected channel
E.g: INIT:RESIstance? CH1
Response: 2.000000\n
Set channel CC/CV priority mode
Command format: [:SOURce][:CVCC]:PRIOrity <channel>,{CC | CV }
Description: Set the CC/CV priority of the selected channel (CC | CV)
E.g: PRIOrity CH1,CC
Description: Set the operation mode to CC priority
Command format: [:SOURce][:CVCC]:PRIOrityq? <channel>
Description: Get the CC/CV priority of the selected channel
E.g: PRIOrityq?
Response: CC\n
Set the measurement average
Command format: SYStem:MEASure:AVERage { LOW| MIDDLE| HIGH}
Description: Set the average measurement value (LOW| MIDDLE| HIGH)
E.g: SYStem:MEASure:AVERage LOW
Command format: SYStem:MEASure:AVERage?
Description: Get average measurement value
E.g: SYStem:MEASure:AVERage?