SPS5000X Series User Manual
WWW.SIGLENT.COM 59
(2) CH1/CH2 parallel
Before realizing serial output, it is necessary to set the mode of SPS and connect the
output port and back-end analog interface.
Mode setting
Setting method: Menu 1.System 5.M/S mode settings to enter the
setting interface, press the knob to display a cursor, turn the knob to select the
configuration of the slave/slave mode, and press the knob to make the setting effective.
Set CH1 to PAR1 according to the above setting method, and the state of CH2 will be
automatically set according to the selection of CH1. After the setting is successful, CH1
is the master, CH2 is the slave, and the setting voltage and current buttons of CH2 cannot
be operated.
Output port connection
-
+
Load 1
-
+
CH3
-
+
CH2
-
+
CH1
SPS5045X
SPS5085X
SPS5165X
Load 2
Analog port connection
User can use the analog connection line option to realize the analog port connection
between the two channels. You can also make your own connection line according to the
corresponding relationship of each port number to realize the connection control between
the two channels.
-
+
Load
-
+
CH2
-
+
CH1
SPS5044X
SPS5084X
SPS5164X