Siglent SPS5083X 80V/45A/1080W Single Channel Switch Mode Power Supply

Manual - Page 84

For SPS5083X.

PDF File Manual, 106 pages, Read Online | Download pdf file

SPS5083X photo
Loading ...
Loading ...
Loading ...
SPS5000X Series User Manual
WWW.SIGLENT.COM 83
*WAI
Command format:
Description: Wait for all outstanding operations to complete before executing any
other commands
E.g:
Measure Command Subsystem
Set voltage value
Command format: [:SOURce]:VOLTage[:SET] <channel>,{<value> | MINimum | MAXimum |DEFault}
Channel={CH1,CH2,CH3}
Description: Set the voltage value of the selected channel
E.g: :SOURce:VOLTage:SET CH1,3
Description: Set the CH1 channel voltage value to 3V
Command format: [:SOURce]:VOLTage[:SET]? <channel>
Description: Get the set voltage value of the selected channel
E.g: :SOURce:VOLTage:SET? CH1
Response: 3.000000\n
Obtain the measured voltage value
Command format: MEASure:VOLTage? <channel>
Description: Get the voltage measurement value of the selected channel
E.g: MEASure:VOLTage? CH1
Response: 2.991442\n
Set current value
Command format: [:SOURce]:CURRent[:SET] <channel>,{<value> | MINimum | MAXimum |DEFault}
Description: Set the current value of the selected channel
E.g: :SOURce:CURRent:SET CH1,2
Description: Set the CH1 channel current value to 3V
Command format: [:SOURce]:CURRent[:SET]? <channel>
Description: Get the set current value of the selected channel
Loading ...
Loading ...
Loading ...