72
Calibration Errors
Code Explanation
702 Cal secured
The calibration is secured.
703 Invalid secure code
An invalid calibration security code was received when attempting to unsecure or
secure the power supply. You must use the same security code to unsecure the power
supply as what was used to secure it, and vice versa. The security code may contain
up to 12 alphanumeric characters.
704 Secure code too long
A security code was received which contained more than 12 characters.
740 Cal checksum failed, secure state
741 Cal checksum failed, string data
745 Cal checksum failed, DAC cal constants
746 Cal checksum failed, readback cal constants
747 Cal checksum failed, GPIB address
748 Cal checksum failed, internal data
750 Cal checksum failed, store/recall data in location 0
751 Cal checksum failed, store/recall data in location 1
752 Cal checksum failed, store/recall data in location 2
753 Cal checksum failed, store/recall data in location 3
754 Cal checksum failed, store/recall data in location 4
755 Cal checksum failed, store/recall data in location 5
756 Cal checksum failed, store/recall data in location 6
Loading ...