51
Bit
Decimal
Val ue
Definition
0(Voltage) 1
The power supply is abnormal in
constant voltage mode.
1(Current) 2
The power supply is abnormal in
constant current mode.
2-3(Not used) 0
Always set to 0.
4(Over-temperature) 16
The fan has a fault condition.
5-8(Not used) 0
Always set to 0.
9(Over Voltage) 512
The power supply is in
over-voltage state.
10-15(Not used) 0
Always set to 0.
The questionable status event register is cleared when:
1. You execute the *CLS (clear status) command.
2. You query the event register with STATus:QUEStionable [:EVENt]? (Status
Questionable Event register) command
The questionable status enable register is cleared when you execute
STATus:QUEStionable:ENABle 0 command.
The Standard Event Register
The standard event register reports the following instrument events: power-on
detected errors, command syntax errors, command execution errors, self-test or
calibration errors, query errors, or an *OPC command. Any or all of these conditions
can be reported in the standard event summary bit of status byte register through the
enable register. To set the enable register mask, you write a decimal value to the
register using the *ESE (Event Status Enable) command.