ARRAY 3672A 35 Volt DC 22.5 Amp Programmable Switching Power Supply

USER’S MANUAL - Page 28

For 3672A. Also, The document are for others ARRAY models: 367*

PDF File Manual, 73 pages, Read Online | Download pdf file

3672A photo
Loading ...
Loading ...
Loading ...
28
Select CV, CC or OV mode calibration by direction key. If “DEF” is selected, all
calibration parameters are restored to default value. CV, CC and OV mode should be
calibrated one by one, which is the correct calibration sequence. Return to calibration
menu after each mode calibration is completed. Press “Clear” key to exist calibration
state.
3.1 CV Mode Calibration
In this mode, three voltage points: 0.5V22V34.5V should be calibrated.
3.1.1 Wiring
3.1.2 0.5V Calibration
Press “Enter” key, you will see the following figure in the display:
Input the value you read from KEITHLEY 2000, which is retained to four decimal
places. Press “Clear” key to remove the wrong input value of current digit. Then press
“Enter” key to confirm.
3.1.3 22 V Calibration
Press “Enter” key to verify and then you will see the followings:
Input the value you read from KEITHLEY 2000, which is retained to four decimal