A suffix for a numeric parameter contained too many characters.
-138 Suffix not allowed
A suffix was received following a numeric parameter which does not accept a suffix.
Example: STAT:QUES:ENAB 18 SEC (SEC is not a valid suffix).
-144 Character data too long
The character data element contained too many characters.
-151 Invalid string data
An invalid character string was received. Check if you have enclosed the character
string in single or double quotes.
Example: DISP:TEXTON
-211 Trigger ignored
A Group Execute Trigger (GET) or *TRG was received but the trigger was ignored.
Make sure that the trigger source is selected to the bus and the trigger subsystem
should be initiated by INIT[:IMM] command.
-213 Init ignored
A Group of INIT[:IMM] was received when the bus line is selected as the trigger
source. Make sure that the trigger source is selected to the bus and the trigger
subsystem is initiated by INIT[:IMM] command, then triggered by *TRG.
-222 Data out of range
A numeric parameter value exceeds the valid range for the command.
Example: TRIG:DEL -3
-223 Too much data
A character string was received but could not be executed because the string length
was more than 40 characters. This error can be generated by the CALibration:STRing
command.
-224 Illegal parameter value
A discrete parameter was received which was not a valid choice for the command.
You may have used an invalid parameter choice.
Example: DISP:STAT ABC (ABC is not a valid choice).
-350 Too many errors
The error queue is full because more than 20 errors have occurred. No additional
errors are stored until you remove errors from the queue. The error queue is cleared
when power has been off, or after a *CLS (clear status) command has been executed.