SDS2000X User Manual 83
10 Address & Data — The oscilloscope will be triggered when the following
conditions are satisfied.
The address’s length must be 10 bits and the address’s value is the same as
the set value.
If you have set either Data1’s or Data2’s value, and the signal has a data is
the same as that value. If you have set both Data1’s and Data2’s value, the
signal should have two consecutive data packets, the first is Data1, and the
second is Data2.
(Note: If the set value is 0xXX, any data value will be matched)
Data Length — When the SDA data length is equal to the value of Byte Length
and address’s length is the same as the set value, the oscilloscope will be
triggered. Byte length is in the range of 1 to 12 bits.
Operation steps:
1. Press Setup to enter the TRIGGER function menu.
2. Press Type and select Serial.
3. Press Protocol and select IIC.
4. Press Trigger Setting softkey.