Pulse Trigger
Trigger on the positive or negative pulse with a specified width.
1. Press the
Trigger Setup button on the front panel to enter the TRIGGER
function menu.
2. Press the Type softkey; turn the Universal Knob to select Pulse and then push
the knob to confirm.
3. Press the Source softkey; turn the Universal Knob to select CH1 or CH2 as the
trigger source.
4. Press the Trigger Level button and turn the Universal Knob to adjust the
trigger level to the desired place.
5. Press the Polarity softkey to select Positive or Negative pulse that to trigger
on. The current trigger polarity is displayed at the upper right corner of the
screen.
6. Press the Limit Range softkey; turn the Universal Knob to select the desired
condition.
<= (less than a time value): trigger when the input signal's positive or
negative pulse time is lower than the specified time value.
For example, for a positive pulse, if you set t (pulse real width) 100ns,
the waveform will trigger.
>= (greater than a time value): trigger when the input signal's positive or
negative pulse time is greater than the specified time value.
For example, for a positive pulse, if you set t (pulse real width) 100ns,
the waveform will trigger.