Siglent SHS820X 200MHz 2CH Handheld Digital Oscilloscope

Manual - Page 189

For SHS820X.

PDF File Manual, 229 pages, Read Online | Download pdf file

SHS820X photo
Loading ...
Loading ...
Loading ...
SHS800X&SHS1000X User Manual
168
WWW.SIGLENT.COM
Do Self Cal
The self-calibration program can quickly make the oscilloscope reach the best
working state to get the most precise measurement values. You can perform
self-calibration at any time, especially when the environment temperature
change is up to or more than 5 ℃. Ensure that the oscilloscope has been warmed
up or operated for more than 30 minutes before the self-calibration.
Do the following steps to do self-calibration:
1. Disconnect all the input channels.
2. Press the Utility button on the front panelïĵŒand then press the Do Self Cal
softkey, and the oscilloscope will pop-out the message box shown as below:
Figure 94 Do Self Cal
3. Press the
F5 button on the front panel to perform the self-calibration
program. During the calibration, most of the keys are disabled.
4. When the self-calibration program is finished, it will pop-out the message
“Press Run/Stop key to exit”. Press the Run/Stop button on the front panel
to exit the calibration interface.