106 User Manual
6.3.2.1 Peak->CF
Set the current peak frequency to the center frequency.
6.3.2.2 Left Peak
Search for and mark the nearest peak which is located at the left side of the current peak and meets the
peak search condition.
6.3.2.3 Right Peak
Search for and mark the nearest peak which is located at the right side of the current peak and meets
the peak search condition.
6.3.2.4 Peak Peak
Execute peak search and minimum search at the same time and mark the results with delta pair markers.
Wherein, the result of peak search is marked with the delta marker and the result of minimum search is
marked with the reference marker.
6.3.2.5 Count Peak
Enable or disable continuous peak search. The default is Off. When enabled, the system will always
execute a peak search automatically after each sweep in order to track the signal under measurement.
6.3.3 Marker->
1. M->CF
Set the center frequency of the analyzer to the frequency of the current marker.
If the Normal marker is selected, the center frequency will be set to the frequency of the current
marker.
If the Delta or Delta Pair marker is selected, the center frequency will be set to the frequency of
the Delta Marker.
2. M->Start Freq
Set the start frequency of the analyzer to the frequency of the current marker.
If the Normal marker is selected, the start frequency will be set to the frequency of the current
marker.
If the Delta or Delta Pair marker is selected, the start frequency will be set to the frequency of the
Delta Marker.
3. M->Stop Freq
Set the stop frequency of the analyzer to the frequency of the current marker.