FPREshoot
Defined as (Vmax-Vhig)/Vamp before the
waveform falling.
Rise Time
Rise Time Measures the time between 10% and 90%
of the first rising edge of the waveform.
Fall Time
Fall Time Measures the time between 90% and 10% of
the first falling edge of the waveform.
BWid
The duration of a burst.
Measured over the entire waveform.
+ Wid
+ Width Measures the time between the first rising
edge and the next falling edge at the waveform 50%
level.
- Wid
-Width Measures the time between the first falling
edge and the next rising edge at the waveform 50%
level.
+ Duty
Measures the first cycle waveform.
Positive Duty Cycle is the ratio between positive pulse
width and period.
Duty
Measures the first cycle waveform. Negative Duty
Cycle is the ratio between negative pulse width and
period.
Phase
The amount one waveform leads or lags another in
time. Expressed in degrees, where 360 degrees
comprise one waveform cycle.
FRR
The time between the first rising edge of source X and
the first rising edge of source Y.
FRF
The time between the first rising edge of source X and
the first falling edge of source Y.
FFR
The time between the first falling edge of source X and
the first rising edge of source Y.
FFF
The time between the first falling edge of source X and
the first falling edge of source Y.
LRR
The time between the first rising edge of source 1 and
the last rising edge of source 2.
LRF
The time between the first rising edge of source X and
the last falling edge of source Y.
LFR
The time between the first falling edge of source X and
the last rising edge of source Y.
LFF
The time between the first falling edge of source X and
the last falling edge of source Y.