SDS1000L User Manual 63
Track mode
Table 2-33 Track mode function menu:
Option Setting Instruction
Cursor Mode Track In this mode, set track cursor measure.
Cursor A
CH1
CH2
NONE
Set the input signal channel that the Cursor
A will measure.
Cursor B
CH1
CH2
NONE
Set the input signal channel that the Cursor
B will measure.
Cur A
Select this option, turn the “Universal” knob
to adjust horizontal coordinate of Cursor A.
Cur B
Select this option, turn the “Universal” knob
to adjust horizontal coordinate of Cursor B.
In this mode, the screen displays two cross cursors. The cross cursor sets the
position on the waveform automatically. You could adjust cursor’s horizontal
position on the waveform by turning the “Universal knob. The oscilloscope
displays the values on the top of the right screen.
To do track cursor measurement, follow these steps:
1. Press CURSOR button to enter the cursor measure function menu.
2. Press the “Cursor Mode” option button to select “Track”.
3. Press the “Cursor A” option button to select the input signal channel.
4. Press the “Cursor B” option button to select the input signal channel.
5. Select “Cur A”, turn the “Universal” knob to move Cursor A
horizontally.
6. Select “Cur B”, turn the “Universal” knob to move Cursor B
horizontally.
7. The measurement values are displayed on the left of the top corner:
AT: The horizontal position of Cursor A (Time cursor centered
around the midpoint of screen).
AV: The Vertical position of Cursor A (Voltage cursor centered
around channel ground level).
BT: The horizontal position of Cursor B (Time cursor centered
around the midpoint of screen).