Siglent SDS1000CNL Series Digital Storage Oscilloscopes

User Manual - Page 36

For SDS1000CNL. Also, The document are for others Siglent models: SDS1000DL, SDS1000CML

PDF File Manual, 135 pages, Read Online | Download pdf file

SDS1000CNL photo
Loading ...
Loading ...
Loading ...
SIGLENT
26 SDS1000L User Manual
If you input a sine signal to channel 1, follow these steps:
1. Measure FFT Amplitude
1) Input a sine signal to channel 1, and press the “AUTO” button.
2) Press the “MATH” button to enter the “MATH” menu.
3) Press the “Operation” option button to select “FFT”.
4) Press the “Source” option button to select “CH1”.
5) Press CH1 button to display CH1 menu.
6) Turn the “S/div” knob to adjust sampling rate (at least double bigger than
frequency of input signal).
7) If FFT display on full screen, press CH1 button again to remove channel
waveform display.
8) Press the “CURSOR” button to enter “Cursor” menu.
9) Press the “Cursor Mode” button to select “Manual”.
10) Press the “Type” option button to select “Voltage”.
11) Press the “Source” option button to select “MATH”.
12) Press the “CurA” option button; turn the “Universal” knob to move Cursor A to
the highest point of the FFT waveform.
13) Press the “CurB” option button, turn the “Universal” knob to move Cursor B to
the lowest point of the FFT waveform.
14) The amplitude (T) displays on the top of the left screen.
Picture 2.6-11
Loading ...
Loading ...
Loading ...