14 SDS1000L User Manual
Table 2-2 Auto set the function item
Function Setting
Acquire Mode Adjusted to Sampling
Display Format Y-T
Display Type Set to Dots for a video signal, set to Vectors
for an FFT spectrum; otherwise, unchanged
Vertical Coupling Adjusted to DC or AC according to the input
signal
Bandwidth Limit Off(full)
V/div Adjusted
VOLTS/DIV adjustability
Coarse
Signal inverted Off
Horizontal position Center
S/div Adjusted
Trigger type Edge
Trigger source Auto detect the channel which has the input
signal
Trigger slope Rising
Trigger mode Auto
Trigger coupling DC
Trigger holdoff Minimum
Trigger level Set to 50%
2.4 Default Setup
The oscilloscope is set up for normal operation when it is shipped from the factory.
This is the default setup. To recall this setup, press the DEFAULT SETUP button.
The options, buttons and controls that change settings when you press the
DEFAULT SETUP button, refer to appendix B.
The DEFAULT SETUP button does not reset the following settings:
● Language option
● Saved reference waveform files
● Saved setup files
● Display contrast
● Calibration data