SIGLENT
52 SDS1000L User Manual
Average: The oscilloscope acquires several waveforms, averages them, and
displays the resulting waveform.
Advantage: You can use this mode to reduce random noise.
Picture 2.9-4 Average Mode
■ Equivalent Time Sampling:
The equivalent time sampling mode can achieve up to 20 ps of horizontal
resolution (equivalent to 50GSa/s). This mode is good for observing repetitive
waveforms.
Real Time Sampling: The scope has the highest Real-time sampling rate
up to 1GSa/s.
■ “RUN/STOP” Button: Press the RUN/STOP button when you want the
oscilloscope to acquire waveforms continuously. Press the button again to stop
the acquisition.
■ “SINGLE” Button: Press the SINGLE button to acquire a single waveform.
Each time you push the SINGLE button, the oscilloscope begins to acquire
another waveform. After the oscilloscope detects a trigger it completes the
acquisition and stops.