Chapter 1 Getting Start
Over compensated
Under Compensated
If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for
the fattest square wave being displayed on the oscilloscope.
Repeat if necessary.
WARNNING: To avoid electric shock while using the probe, be sure the perfection of the
insulated cable, and do not touch the metallic portions of the probe head while it is connected
with a voltage source.
1.6 Self Calibration
The self calibration routine lets you optimize the oscilloscope signal path for maximum
measurement accuracy. You can run the routine at any time but you should always run the
routine if the ambient temperature changes by 5v or more. For accurate calibration, power on
the oscilloscope and wait twenty minutes to ensure it is warmed up. To compensate the
signal path, disconnect any probes or cables from the input connectors. Then, access the
“Utility -> Calibration” option and follow the directions on the screen. The self calibration
routine takes about several minutes.
Hantek6102BE/Hantek6212BE Users Manual 13