User Manual Braun NTF3000 No touch forehead thermometer

Braun NTF3000 No touch forehead thermometer - Use Manual - Use Guide PDF.
Documents: Go to download!
  • Owner's manual - (English, French, Portuguese, Spanish, German, Danish, Dutch - Holland, Finland, Greek, Italian, Norway, Swedish)



User manual No touch forehead thermometer

Product description

1. Power button

2. Scanner

3. LCD display

4. Temperature button

5. Battery cover

6. Protective scanner cap

7. Silent mode switch

8. Guidance light

WARNINGS AND PRECAUTIONS

  • Keep out of reach of children under 12 years.
  • Never use the thermometer for purposes other than those it has been intended for. Please follow the general safety precautions when using on children.
  • Never immerse the thermometer into water or other liquids (not waterproof). For cleaning and disinfecting please follow the instructions in the “Care and cleaning” section.
  • Do not store this thermometer in temperature extremes below -25 °C or over 60 °C (below -13 °F or over 140 °F) or in excessive humidity (above 95% non-condensing relative humidity).
  • If thermometer is stored in a location that is cooler or warmer than where it is being used, let it sit in the patient’s room for 10 minutes before taking the measurement.
  • Do not use the thermometer if there are signs of damage on the scanner or on the thermometer itself. If damaged, do not attempt to repair the product.
  • Never insert a sharp object into the scanner area or any other open surface on the thermometer.
  • This thermometer consists of high-quality precision parts. Do not drop the instrument. Protect it from severe impact and shock. Do not twist the instrument or the measuring sensor.
  • This thermometer is intended for household use only.
  • Use of this thermometer is not intended as a substitute for consultation with your physician.
  • Temperature elevation may signal a serious illness, especially in neonates and infants, or in adults who are old, frail, or have a weakened immune system. Please seek professional advice immediately when there is a temperature elevation and if you are taking temperature on:
    • Neonates and infants under 3 months (Consult your physician immediately if the temperature exceeds 37.4 °C [99.4 °F])
    • Patients over 60 years of age (Fever may be blunted or absent in older patients)
    • Patients having diabetes mellitus or a weakened immune system (e.g., HIV positive, cancer chemotherapy, chronic steroid treatment, splenectomy)
    • Patients who are bedridden (e.g., nursing home patient, stroke, chronic illness, recovering from surgery)
    • A transplant patient (e.g., liver, heart, lung, kidney).
  • This thermometer is not intended for pre-term babies or small-for-gestational age babies.
  • This thermometer is not intended to interpret hypothermic temperatures.
  • Do not allow children to take their temperatures unattended.
  • Please consult your doctor if you see symptoms such as unexplained irritability, vomiting, diarrhea, dehydration, changes in appetite or activity, seizure, muscle pain, shivering, stiff neck, pain when urinating, etc., even in the absence of fever.
  • Even in the absence of fever, those who exhibit a normal temperature may still need to receive medical attention. People who are on antibiotics, analgesics, or antipyretics should not be assessed solely on temperature readings to determine the severity of their illness.
  • Do not modify this equipment without the authorization of the manufacturer.

Why Braun No touch + forehead thermometer?

Measurement in under 2 seconds

  • The innovative infrared technology allows measurement of forehead temperature in under 2 seconds whether you use the touch or no-touch option.

Accurate and reliable

  • Due to the unique patented technology, the No touch + forehead thermometer captures the heat naturally given off by the forehead to calculate body temperature value. Get the same professional accuracy whether touching your child’s forehead or holding up to 5 cm away.

Easy to use

  • The No touch + forehead thermometer is non-invasive. A measurement can be taken even while a child is sleeping.
  • The No touch + forehead thermometer is less threatening to a child than a rectal thermometer and easier to use than other methods.

Safe and hygienic

  • No-touch option helps minimize spreading of germs.
  • Completely safe for use on children and adults.

How does Braun No touch + forehead work?

  • The No touch + forehead thermometer measures infrared energy radiated from the skin at the center of the forehead area. This captured energy (which is twice as much thermal energy compared to a traditional forehead thermometer1 ) is collected through the lens and converted to a body temperature value.
  • The No touch + forehead thermometer has been clinically tested and proven to be safe and accurate when used in accordance with its operating instruction manual.
  • A traditional forehead product without any optical system to capture radiated heat.
  • Fever guidance feature: Fever guidance helps you to better understand the meaning of your child’s temperature with the color indicated on the display. The screen displays green if no fever, yellow when the temperature is elevated and a red alert for a possibly high fever. 10 audible beeps indicate temperatures above 37.4 °C (99.4 °F) to alert the user that the patient may have a fever.

  • NOTE: Backlight remains OFF when below temperature readings are registered: 34.4° ≤ T ≤ 35.7 °C (93.9° ≤ T ≤ 96.3 °F)

How to use your Braun No touch + forehead

1 Remove cap

2 Power on

  • Press and release the power button once. Backlight will come on and the start-up sequence begins.
  • NOTE: Make sure to remove the protective scanner cap before taking a measurement.
  • When the thermometer is ready and correctly positioned, a horizontal line of dashes (“---”) will appear on the screen.

3 Position

  • Position thermometer on or up to 5 cm away from the center of the forehead, just between the eyebrows.
  • For No touch readings, the yellow guidance light will show you where you are aiming. If the eyebrow area is covered with hair, sweat or dirt, please clean the area beforehand to improve the reading accuracy. It is important to hold the thermometer and the forehead steady during measurement. Movement will impact the temperature reading.

4 Take temperature

  • Press the temperature button (you can press and hold the button or press and release it). When the thermometer is placed correctly, the screen displays a dashed line animation while it takes a reading. After the animation sequence (under 2 seconds), the display shows the temperature reading. The appropriate fever light color is displayed on screen and the confi rmation beep is heard.
  • If the thermometer is positioned too far away from the forehead, it will prompt you to move the thermometer closer by displaying diagram and “0-5 cm”. Slowly move the thermometer toward the forehead until the horizontal line animation starts and the temperature reading is displayed.

5 Read temperature

  • Remove the thermometer and read the temperature. For low (no backlight) and acceptable (green backlight) temperature range, you will hear a single long beep for 2 seconds. For elevated (yellow backlight) and possible high fever (red backlight) temperature range, you will hear 10 short beeps.
  • To repeat: Go to step 3.

6 To turn off Press the power button to turn off . Device will also shut off automatically after 60 seconds of no use

Temperature taking hints

  • It is important to know each individual’s normal temperature when they are well. This is the only way to accurately diagnose a fever. Take multiple readings when healthy to determine normal temperature.
  • A child’s normal temperature can be as high as 37.7 °C (99.9 °F) or as low as 36.1 °C (97.0 °F). Be sure to note this unit reads 0.5 °C (0.9 °F) lower than a rectal digital measurement.
  • Patient must be inside for 30 minutes before taking a measurement.
  • Note: Patient and the thermometer should be in the same ambient temperature for at least 10 minutes.
  • Always hold the thermometer and the forehead steady when taking a reading. Do not move the thermometer until you hear the fi nal beep. Don’t take a measurement while or immediately after nursing a baby.
  • Patients should not drink, eat, or be physically active before/while taking the measurement. Remove hats and wait 10 minutes before taking a measurement.
  • Before taking a measurement, remove dirt or hair from the forehead area. Wait 10 minutes after cleaning before taking measurement.
  • Always take the temperature exactly as directed. Temperature results may vary if positioned in the wrong location.
  • For patients measuring their own temperature, it is recommended to use the “touch” option instead of “no touch.

In the following situations it is recommended that three temperatures in the same location be taken and the highest one taken as the reading:

  • Newborn infants in the fi rst 100 days.
  • Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical.
  • When the user is learning how to use the thermometer for the fi rst time until he/she has familiarized himself/herself with the instrument and obtains consistent readings.

General precautions:

  • Remove any sweat prior to measuring with a dry cloth.
  • Avoid any cooling or warming cloths on the forehead for at least 30 minutes prior to measurement.
  • Do not take temperature measurements over scar tissue, open sores or abrasions.
  • Keep the measurement sensor and lens clean. Avoid directly touching the sensor or lens.
  • Always store the thermometer with the protective cap in place to prevent dirt and scratches from damaging the device.
  • Make sure to remove the protective cap before taking a measurement and to put the cap back on after using the thermometer.

Changing the temperature scale

  1. Open battery compartment and remove the batteries.
  2. °C / °F switch is accessible in the center of the battery compartment.
  3. Slide switch to °C or °F to set preferred temperature scale.
  4. Replace the batteries and close the battery door.

Changing the sound mode

  • The Braun No touch + forehead thermometer allows you to silence the beeps on the thermometer to avoid waking a sleeping child. To activate the silence feature, slide the switch to the silent mode position and take a temperature reading. An icon will appear on the screen to indicate that the thermometer is in silent mode.

Calibration

  • This device was designed and manufactured for a long service life, however it is generally recommended to have the device inspected once a year by the Authorized Service Center located in your country to ensure correct function and accuracy.
  • Note: The accuracy checking is not a free service and therefore we recommend that you contact the Authorized Service Center to get a quotation before you send out the product.
  • This device is not intended as a substitute for regular check-ups by your doctor, please consult your doctor if you have any doubt about the temperature reading.
  • LOT and Serial numbers of your product must be provided as they are essential to record and follow up your inquiry or claim. The LOT and Serial numbers are located in the battery compartment.
  • The production date is given by the LOT number and can be deciphered as explained below: The fi rst 3 numeric digits in the LOT number represent the day of the year of manufacture. The next 2 numeric digits represent the last two numbers of the calendar year of manufacture and the letter(s) at the end designate the manufacturer of the product. (E.g.: LOT No.: 12313tav this product was made on the day 123, year 2013 at the manufacturer code tav.)

Errors and troubleshooting

1. When 20% of battery life is left, the display fl ashes the low battery warning symbol; however, the device can still work until the battery life has 0% left. --> Replace batteries.

2. If the steady battery icon is the only symbol shown on the display, the device cannot work. The battery should be replaced immediately --> Replace batteries.

3. This message displays when the measured temperature is lower than 34.4 °C (93.9 °F) or higher than 42.2 °C (108 °F) or when the room temperature is outside the operating range of 15 °C – 40 °C (59 °F – 104 °F). --> Re-measure the temperature, carefully following the instructions in “How to use” section.

4. Blank display. Thermometer does not have power. --> Please check if the batteries have been loaded correctly. Also check polarity ( and ) of batteries. Contact customer service if thermometer still does not function.

Care and cleaning

  • Use an alcohol swab or cotton swab moistened with alcohol (70% isopropyl) to clean the thermometer casing and the measuring probe. Ensure that no liquid enters the interior of the thermometer.
  • Never use abrasive cleaning agents, thinners or benzene for cleaning and never immerse the instrument in water or other cleaning liquids. Wait 10 minutes after cleaning before taking a temperature measurement. Make sure to replace the protective cap after use to prevent scratches or damage from occurring to the thermometer. Never insert a sharp object into the scanner area or any other open surface on the thermometer.

Replacing the batteries

  • The No touch + forehead thermometer comes with 2 AA batteries. Replace with 2 new AA batteries when the fl ashing battery symbol appears on the LCD display. To change the batteries, slide open the battery cover and remove batteries. Replace the batteries being sure to align properly as indicated inside the battery compartment. Remove the battery from the product if it is not required for extended periods of time in order to avoid damage to the thermometer resulting from a leaking battery.
  • To protect the environment, dispose of empty batteries at appropriate collection sites according to national or local regulations.

Product specifi cations

  • Type: No touch + forehead thermometer (NTF 3000)
  • Measuring range: 34.4 °C – 42.2 °C (93.9 °F – 108 °F)
  • Resolution: 0.1 °C (0.1 °F)
  • Laboratory Accuracy: ±0.2 °C for the range 35 °C – 42 °C (±0.4 °F for 95 °F – 107.6 °F) ±0.3 °C (outside this temperature range) (Ambient Temperature: 15 °C to 40 °C (59 °F to 104 °F))
  • Display: Liquid Crystal Display, 4 digits plus special icons
  • Acoustic: Audio: Normal temp range = single long beep for 2 sec duration Fever = Red or Yellow temp range: 10 short beeps Operating temperature: 15 °C – 40 °C (59 °F to 104 °F) Automatic Switch-off : Approx. 60 seconds after last measurement has been taken
  • Weight: 99.5 g (with batteries), 77.1 g (w/o batteries)
  • Long term storage ranges
  • Storage/transport temperature: -25 °C to 60 °C (-13 °F – 140 °F)
  • Humidity: 15–95% non-condensing
  • Battery: (2) AA Batteries - at least 1000 measurements
  • Pressure: 700-1060 hPA (0,7-1,06 atm

This product conforms with the provisions of the European directive 93/42/EEC (Medical Device Directive). This infrared thermometer meets requirements established in ASTM Standard E 1965-98 for the thermometer system. Full responsibility for the conformance of the product to the standard is assumed by Kaz Europe Sàrl, Place Chauderon 18, CH-1003 Lausanne, Switzerland.

ASTM laboratory accuracy requirements for the thermometer only in the display range of 36 °C to 39°C (96.8 °F to 102.2 °F) for infrared thermometers is ±0.3 °C (±0.5 °F), whereas for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ±0.1 °C (±0.2 °F).

This device conforms to the following standards:

  • EN 60601-1:2006 Medical electrical equipment. General requirements for basic safety and essential performance
  • EN 80601-2-56 Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement ASTM 1965E – ASTM Standard for Infrared Clinical Thermometers
  • EN 12470-5: 2003 «Clinical thermometers» – Part 5: Performance of infrared ear thermometers (with maximum device)
  • EN 60601-1-2: Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic compatibility – Requirement and tests
  • NOTE: Do not use this device in the presence of electromagnetic or other interference outside the normal range specifi ed in EN 60601-1-2.
  • EN 980: 2008 Symbols for use in labeling of medical devices
  • EN 1041: 2008 Information supplied by the manufacturer of medical devices
  • EN 60601-1-11: 2010 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment


See other models: IRT3030 IRT6020 IRT6520 MQ505 MQ30BK

Braun NTF3000 | File type: PDF | Filename: Braun No touch + forehead thermometer – NTF3000.pdf | Size: 7.03 MB | Language: English, French, Portuguese, Spanish, German, Danish, Dutch - Holland, Finland, Greek, Italian, Norway, Swedish

Download User Manual

Product Infomation and Reviews - NTF3000 photo

NTF3000 Photo

Braun in United States

1 (888) 977-3578 Monday through Friday from 9:00 am – 8:00 pm (ET).

Braun NTF3000 Questions and Answers

Find answers or Ask a question.
  • Total questions: 1
  • Questions unAnswered : 1

Q: what does a red error indication mean and how do you clear it? Reply