Loading ...
Loading ...
Loading ...
95
Recommended separation distances
Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT
or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFESUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the Blood pressure moni-
tor (SBC 53).
The Blood pressure monitor (SBC 53) is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Blood pressure monitor (SBC 53) can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Blood pres-
sure monitor (SBC 53) as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter (m)
80 MHz to 800 MHz 800 MHz to 2.5 GHz
0,01 0.1167 0.2334
0,1 0.3689 0.7378
1 1.1667 2.3334
10 3.6893 7.3786
100 11.6667 23.3334
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
Loading ...