34
In such cases, repeat the measurement and/or data transfer.
Ensure that the cuff line is correctly attached and that you do
not move or speak during the measurement.
9. Error messages/troubleshooting
In the event of errors, the error message
_
appears on the
display. Error messages may appear if:
it was not possible to record the pulse:
1
;
you move or speak during the measurement:
2
;
the cuff is fastened too tightly or loosely:
3
;
errors occur during the measurement:
4
;
the pump pressure is higher than 300 mmHg:
5
;
the batteries are almost empty :
6
;
the data could not be sent via Bluetooth
®
.
The word mark Bluetooth
®
and accompanying logo are registe-
red trademarks of Bluetooth
®
SIG, Inc. Any and all use of these
marks by Hans Dinslage GmbH is done so under licence. Other
trademarks and trade names are the property of the relevant
holder.
10. Technical specifications
Model No. SBC 53
Measurement
method
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Measurement range Cuff pressure 0 – 300mmHg,
Systolic 50 – 250mmHg,
Diastolic 30 – 200mmHg,
Pulse 40 – 180 beats/minute
Display accuracy Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
curacy
Max. permissible standard deviation
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory 2 x 60 memory spaces
Dimensions L 78 mm x W 63 mm x H 23 mm
Weight 66.5 g (Without cuff and batteries)
Cuff size 14 to 19.5 cm
Permissible operating
conditions
+ 5°C to + 40°C, 15-93% relative humid-
ity (non-condensing)
Permissible storage
conditions
-25°C to + 70°C, ≤ 93% relative humidity,
700 –1060 hPa ambient pressure
Power supply 2x 1.5V AAA batteries