Guidance and manufacturer’s declaration – electromagnetic immunity for EQUIPMENT and SYSTEMS that are not LIFE-
SUPPORTING
The Blood pressure monitor (SBC 53) is intended for use in the electromagnetic environment specified below. The customer
or the user of Blood pressure monitor (SBC 53) should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the Blood pressure monitor (SBC 53), inclu-
ding cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance.
Conducted RF
IEC 61000-4-6
3 V
rms
150 kHz to
80 MHz
3 V
rms
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
80 MHz to 800 MHz
800 MHz to 2.5 GHz