Directions For Use 15
Further information
The level of blood pressure is subject to fluctuations, even with
healthy people. It is important to understand that comparable
measurements always require the same conditions. If, in
spite of observing all these factors, the fluctuations are larger
than 15 mmHg and/or you hear irregular pulse tones on several
occasions, consult your doctor.
This device has been subjected to strict clinical tests. The
conformity of this product with the requirements of the CE-
guideline 93/42 EWG, council from June 14, 1993 concerning
Medical Products and Medical Product-Legislation from August
2, 1994, was established through a conformity check process.
Consult your dealer if there are technical problems with the blood
pressure monitor. Never attempt to repair the instrument
The device frequently fails to measure the
blood pressure values, or the values
measured are too low or too high.
1. Fit the cuff correctly on the arm.
2. Before starting the measurement,
make sure the cuff is not fitted too
tightly and that there is no rolled-up
sleeve exerting pressure on the arm
above the measuring position. Take off
articles of clothing if necessary.
3. Measure the blood pressure again in
complete silence.
Every measurement results in different
values, although the device functions
normally and normal values are displayed.
Read the following information and the
points listed under «Common sources of
error». Repeat the measurement.
The blood pressure values measured differ
from those measured by the doctor.
Record the daily development of the
values and consult your doctor about
them.
After the instrument has inflated the cuff
the pressure falls very slowly or not at all
(no reasonable measurement possible).
Missing plastic ring (distance piece) at
tube connection. This may only be
removed for servicing reasons. Push this
ring over the tube connector so it fits
properly and repeat the measurement.
Malfunction Remedy