FED (E2) 03/2019 page 42/57
12.4 FED technical data
Chamber size
400
720
Exterior dimensions
Width, net
mm / inch
1234 / 48.58
1234 / 48.58
Height, gross (incl. feet/castors)
mm / inch
1022 / 40.24
1528 / 60.16
Depth
mm / inch
765 / 30.12
865 / 34.06
Depth gross (incl. door handle, and exhaust duct)
mm / inch
855 / 33.66
955 / 37.60
Wall clearance rear (minimum)
mm / inch
100 / 3.94
100 / 3.94
Wall clearance side (minimum)
mm / inch
160 / 6.30
160 / 6.30
Exhaust duct, outer diameter
mm / inch
52 / 2.05
52 / 2.05
Doors
Number of doors
2
2
Interior dimensions
Width
mm / inch
1000 / 39.37
1000 / 39.37
Height
mm / inch
800 / 31.50
1200 / 47.24
Depth
mm / inch
500 / 19.69
600 / 23.62
Interior volume
l / cu.ft.
400 / 14.3
720 / 25.7
Steam space volume
l / cu.ft.
498 / 17.60
869 / 30.71
Racks
Quantity of racks (regular)
2
2
Quantity of racks (max.)
10
15
Maximum load per rack
Kg / lbs
35 / 77
45 / 99
Permitted total load
Kg / lbs
90 / 199
120 / 265
Weight
Weight (empty)
Kg / lbs
145 / 320
195 / 430
Temperature data
Temperature range, 5 °C / 9 °F above ambient up
to
°C / °F 300 / 572 300 / 572
Temperature fluctuation
at 70 °C / 158 °F
± K
0.1
0.1
at 150 °C / 302 °F
± K
0,7
0,3
at 300 °C / 572 °F
± K
1.4
0.9
Temperature uniformity
(variation)
at 70 °C / 158 °F
± K
0.8
0.8
at 150 °C / 302 °F
± K
3,8
3
at 300 °C / 572 °F
± K
11
8.5
Heating up time
to 70 °C / 158 °F
min
15
8
to 150 °C / 302 °F
min
29
29
to 300 °C / 572 °F
min
86
75
Recovery time after door
was opened for 30 sec
at 70 °C / 158 °F
min
5
8
at 150 °C / 302 °F
min
6
8
at 300 °C / 572 °F
min
15
18
Ventilation data
Air change
at 70 °C / 158 °F
x/h
17
11
at 150 °C / 302 °F
x/h
18
12
at 300 °C / 572 °F
x/h
16
10
Loading ...