FED (E2) 03/2019 page 13/57
If you need to return the chamber, please use the original packing and observe the guidelines for safe
lifting and transportation (chap. 3.2).
For disposal of the transport packing, see chap. 10.1.
Note on second-hand chambers (Ex-Demo chambers):
Second-hand chambers have been used for a short time for tests or exhibitions. They are thoroughly
tested before resale. BINDER ensures that the chamber is technically sound and will work flawlessly.
Second-hand chambers are marked with a sticker on the chamber door. Please remove the sticker before
commissioning the chamber.
3.2 Guidelines for safe lifting and transportation
The front castors of chambers size 720 can be blocked by brakes. Please move the chambers with
castors only when empty and on an even surface, otherwise the castors may be damaged. After operation
please observe the guidelines for temporarily decommissioning the chamber (chap. 10.2).
CAUTION
Sliding or tilting the chamber.
Damage to the chamber.
Risk of injury by lifting heavy loads.
Transport the chamber only in its original packaging.
Secure the chamber with transport straps for transport.
Do NOT lift or transport the chamber using the door handle or the door.
Do NOT lift chambers by hand.
Place chambers using technical devices (fork lifter) on the transport pallet. Set the fork
lifter only from the rear in the middle of the chamber. Make sure to place all the lateral
supports of the chamber on the forks.
Transport chambers ONLY with the original transport pallet. Set the fork lifter only to
the pallet. Without the pallet the chamber is in imminent danger of overturning!!
Permissible ambient temperature range during transport: -10 °C to +60 °C.
You can order transport packing and pallets for transportation purposes from BINDER Service.
3.3 Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 10.2).
Permissible ambient temperature range during storage: -10 °C to +60 °C.
Permissible ambient humidity: max. 70 % r.H., non-condensing
When after storage in a cold location you transfer the chamber to its warmer installation site,
condensation may form. Before start-up, wait at least one hour until the chamber has attained ambient
temperature and is completely dry.