Binder FED 400

User Manual - Page 26

For FED 400.

PDF File Manual, 57 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
FED (E2) 03/2019 page 26/57
6.4.5 Setting the interface mode and, if appropriate, the printer interval
1. Press down button for approx. 5 seconds.
The display alternately shows unitand the temperature unit:
2. Press again button
.
The display alternately shows rASdand the set-point gradient:
3. Press again button
.
The display alternately shows Adrand the chamber address:
4. Press again button
The display alternately shows tFCtand the timer function:
5. Press again button
The display alternately showsPFCt and the actual setting of the interface mode:
6. Set the desired interface mode with buttons
.
Settings: Modbus = “0” printer = “1”
In case of temperature data acquisition by the APT-COM™ 4 Multi Management Software
(option, chap. 8.1) interface mode ”0” (Modbus) must be selected.
The setting is automatically adopted after 2 seconds.
If interface mode “1” (printer) has been selected, the printer interval for the automatic output can be set in
an additional menu step:
7. Press again button
The display alternately shows Prtand in the entry level the actual setting of the printer interval: