KBF / KBF-UL + KMF (E6) 04/2020 page 26/163
With temperatures below +3 °C / 37.4 °F, water must be completely removed from the humidifying sys-
tem.
NOTICE
Danger of freezing in the steam generator when storing the chamber below +3 °C /
37.4 °F with filled steam humidifying system.
Damage to the chamber.
Contact BINDER Service before any storage below +3 °C / 37.4 °F.
Permissible ambient humidity: max. 70 % r.h., non-condensing.
After extensive operation at humidity levels > 70% r.h., condensation from excessive humidity can lead to
corrosion during storage. In this case the chamber must first be dried.
NOTICE
Danger of corrosion on the housing due to condensation by excess humidity after
operating at humidity values > 70 % r.h. for a long period.
Damage to the chamber.
Let the chamber dry for several days before shut-down:
Set the humidity to 0 % r.h. To enable dehumidification, the humidifying and dehu-
midifying system must be activated (deactivated operation line “Humidity off”, chap.
7.3 and setting “Control on”, chap. 6.3).
Set the temperature set point to 60 °C / 140 °F for approx. 2 hours (Manual mode).
Only then, shut down the chamber at the main power switch (1) and close the tap of
the water supply.
When after storage in a cold location you transfer the chamber to its warmer installation site, condensa-
tion may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature
and is completely dry.
In case of a prolonged temporal decommissioning: Leave the chamber door open or remove the access
port plugs.
3.4 Location of installation and ambient conditions
Set up the constant climate chamber on a flat, even surface, and in a well-ventilated, dry location and
align it using a spirit level. The site of installation must be capable of supporting the chamber’s weight
(see technical data, chap. 23.5). The chambers are designed for setting up inside a building (indoor use).
NOTICE
Danger of overheating due to lack of ventilation.
Damage to the chamber.
Do NOT install the chamber in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
Observe the prescribed minimum distances when installing the chamber.
Do not install or operate the chamber in potentially explosive areas.
DANGER
Danger of explosion due to combustible dusts or explosive mixtures in the vicinity
of the chamber.
Serious injury or death from burns and / or explosion pressure.
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.