KBF / KBF-UL + KMF (E6) 04/2020 page 119/163
In case of power failure, transmission of the alarm via zero-voltage relay outputs remains active for the
duration of the power failure. Afterwards, both contacts will close automatically.
When using the APT-COM™ 4 Multi Management Software (option, chap. 19.1) via the inter-
face of the constant climate chamber for data acquisition, the alarm is not automatically
transmitted to the APT-COM™ protocol.
Set the tolerance limits for recording limit value excesses separately in APT-COM™ 4.
19.6 Water protected internal socket (option for KBF)
The internal socket is splash proof.
IP system of protection 67 230 V 1N ~ 50-60 Hz
Charge max. 500 W
The maximum permitted operating temperature with this option is 90 °C / 194 °F.
DANGER
Electrical hazard and danger of damages when exceeding the permitted maximum
temperature.
Deadly electric shock. Damage to the internal socket.
Do NOT exceed the temperature set-point of 90 °C / 194 °F.
Set the safety controller to 90 °C / 194 °F.
With optional temperature safety device class 3.3: Set the mechanical thermostat class
3.1 to 90 °C / 194 °F.
Heat emission of electrical devices connected inside the chamber may modify the temperature
and humidity range.
NOTICE
Risk of short circuit caused by penetration of moisture into the socket.
Damage to the chamber.
Use the supplied plug only (IP protection type 67). Plug it in and tighten it by screwing
to secure contact.
If the socket is not used, close the screw lid and turn it to secure.