Binder MKT 720 dynamic climate chamber with expanded low-temperature range

User Manual - Page 57

For MKT 720.

PDF File Manual, 163 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
MK / MKT (E5) 08/2020 page 57/162
“Ramp” and “Step” mode example (representation of a temperature course)
W/°C
t/min.
0
20
40
60
80
100
1
0
30 40 60 70 100
130
1
2
3 4 5 6 7
8
Corresponding program table
Section No.
Duration
[hh:mm:ss]
Temperature
[°C]
Fan (MK 56)
[%]
Ramp or Step
1
00:10:00
40.0
xxxx
Step
2
00:20:00
60.0
xxxx
Step
3
00:10:00
80.0
xxxx
Step
4
00:20:00
40.0
xxxx
Step
5
00:10:00
40.0
xxxx
Ramp
6
00:30:00
80.0
xxxx
Ramp
7
00:30:00
80.0
xxxx
Ramp
8
00:00:01
20.0
xxxx
Ramp
9.7.3 Special controller functions via operation lines
You can define the switching state of up to 16 operation lines (control contacts). They are used to activate
/ deactivate special controller functions.
Operation line Idle modeactivates / deactivates the operating mode Idle mode”.
Operation lines Switching output 1“ up to „4“ can be used to turn on and off any equipment connected
to the zero-voltage switching outputs (DIN sockets (7) and (8) (option, chap. 19).
Operation line Compr. air dryerserves to activate the compressed air dryer (option, chap. 20.6).
Operation line “Compressed air valve” serves to open the solenoid valve of the compressed air connec-
tion (for options compressed air connection, chap. 20.5, or compressed air dryer, chap. 20.6)
Operation line Anti-condensation protectionserves to switch the anti-condensation protection (chap.
18).
The other operation lines are without function.