MK / MKT (E5) 08/2020 page 26/162
NOTICE
Danger by stacking.
Damage to the chambers.
Do NOT place the chambers on top of each other.
Other requirements
To completely separate the chamber from the power supply, you must disconnect the power plug. Install
the chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger.
With an increased amount of dust in the ambient air, clean the condenser fan several times a year. We
recommend checking the fan grid (behind the left maintenance access flap) every week. In case of visible
dirt accumulation, disconnect the chamber and clean the fan grid by suction.
Avoid any conductive dust in the ambiance according to the chamber layout complying with pollution degree
2 (IEC 61010-1).
4. Installation and connections
4.1 Connection of cooling water outlet for water cooling (option for MK 56)
An enclosure inside the chamber contains the connection kit for the cooling water inlet and outlet.
Fasten the cooling hose to the connection OUT” (16) on the rear of the chamber (screw thread ¾’’).
You can use a part of the supplied tap water hose as a drainage hose. In case another hose is used, it
has to be permanently resistant against max. 50 °C / 122 °F.
Protect both ends of the drainage hose with two of the four supplied hose clamps. Before turning on the
chamber, check the connection for leaks.
4.2 Connection of cooling water inlet for water cooling (option for MK 56)
Connect the cooling water outlet before connecting the cooling water inlet.
Type of suitable water quality:
Water intake temperature: max. 10 °C / 50 °F.
pH value 4-7
connection pressure: 4 to 10 bar
BINDER GmbH is NOT responsible for the water quality at the user’s site.
Any problems and malfunctions that might arise following use of water of deviating quality is
excluded from liability by BINDER GmbH.
The warranty becomes void in the event of use of water of deviating quality.
Connection:
An enclosure inside the chamber contains the connection kit for the cooling water inlet and outlet.
Fasten the cooling water hose to the connection IN” (17) on the rear of the chamber (screw thread ¾’’).
Install the water supply connection using either the enclosed water hose or another pressure-resistant
one. To accomplish this, remove the cover of the freshwater connection IN” (17) on the rear of the
chamber.
The nominal diameter of the supplied water hose is ½ ", the length is 3m. The hose can be halved for
the inlet and the outlet.