8. Timer program: stopwatch function
During an entered duration the controller constantly equilibrates to the setpoints entered in Fixed value
operation mode (temperature, fan speed (with MK 56), configuration of the operation lines). This duration
can be entered as a “Timer program”. During the program runtime, any setpoint changes do not become
effective; the controller equilibrates to the values which were active during program start.
8.1 Starting a timer program
In Normal display press the Program start icon to access the “Program start” menu.
“Program start” menu
In the field Program typeselect Timer program”.
Select the field Program durationand enter the desired program duration. Press the Confirm icon.
Select the field Program startand enter the desired start time of the program in the Program start
entry menu. Press the Confirm icon. The program delay time until program start begins.
Normal display.
Information on the bottom of the screen indicates the
currently running program and the time already
passed. The grey bar shows how much time of the
whole time is elapsed.
8.1.1 Performance during program delay time
During the configured program delay time until program start, the controller equilibrates to the current set-
points of Fixed value operation mode. Modifications of these setpoints are possible but become effective
only after the timer program is finished. When the configured moment for program start is reached, the
program delay time ends and the program starts running. The controller equilibrates to the values which
had been active during program start