MK / MKT (E5) 08/2020 page 28/162
For the installation of the voltage and frequency changer behind the chamber, provide a rear wall distance
of the chamber of approx. 1 m / 3.3 ft.
If possible, fix the voltage and frequency changer at the chamber. For this purpose, an Allen key size 4 is
required. Connect the slots at the end of the chassis with two M6 screws to the threads provided below on
the rear panel of the chamber (see Figure 12).
NOTICE
Danger of overheating due to lack of ventilation.
Damage to the voltage and frequency changer.
Do NOT install the voltage and frequency changer in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
The voltage and frequency changer is equipped with four castors. The rear castors can be easily locked
via the attached brakes
4.4 Electrical connection
4.4.1 Information on connecting the alternating climate chamber
The chambers are supplied ready for connection.
MK 56: The chambers come with an IEC connector plug and are equipped with 1 internal overload release
against excess-current.
MK/MKT 115,240,720: The chambers come with a fixed power connection cable of at least 1800 mm /
70.87 in in length. They are equipped with three internal overload releases against excess-current.
Model Power plug
Nominal voltage +/- 10% at the
indicated power frequency
Current type Chamber fuse
MK 56
(230V)
IEC connector plug
(grounded plug)
230 V at 50 Hz 1 N~
16 Amp
internal
MK 56
(240V)
IEC connector plug
(grounded plug)
240 V at 60 Hz 2~
16 Amp
internal
MK 115
MKT 115
MK 240
MKT 240
MK 720
CEE plug 5-poles
16 Amp
400 V at 50 Hz
3 N~
16 Amp
3 x internal
MKT 720
CEE plug 5-poles
32 Amp
400 V at 50 Hz 3 N~
25 Amp
3 x internal
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chambers protective conductor meets the latest
technology. The protective conductors of the socket and plug must be compatible!
DANGER
Electrical hazard due to missing protective conductor connection.
Deadly electric shock.
Make sure that the chamber’s power plug and the power socket match and se-
curely connect the electrical protective conductors of the chamber and the house
installation.