MK / MKT (E5) 08/2020 page 25/162
NOTICE
Danger of overheating due to lack of ventilation.
Damage to the chamber.
Do NOT install the chamber in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
Observe the prescribed minimum distances when installing the chamber.
Do not install or operate the chamber in potentially explosive areas.
DANGER
Danger of explosion due to combustible dusts or explosive mixtures in the vicinity
of the chamber.
Serious injury or death from burns and / or explosion pressure.
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.
Ambient conditions
Permissible ambient temperature range during operation: +18 °C / 64.4 °F to +32 °C / 89.6 °F. At ele-
vated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +25 °C / 77 °F to which the specified technical data relate. For other ambient con-
ditions, deviations from the indicated data are possible.
Permissible ambient humidity: 70 % r.h. max., non-condensing.
When operating the chamber at temperature set-points below ambient temperature, high ambient humidity
may lead to condensation on the chamber.
Installation height: max. 2000 m / 6562 ft. above sea level.
Minimum distances:
Distance between each chamber when placing several chambers of the same size side by side: 250
mm / 9.84 in
Lateral wall distance: 200 mm / 7.87 in
Wall distance rear: 300 mm / 11.81 in
Chambers with optional water cooling (without options compressed air dryer and / or voltage and fre-
quency changer): Wall distance rear 100 mm / 3.94 in.
Chambers with optional compressed air dryer: Wall distance rear approx. 1 m / 3.28 ft so that it is pos-
sible to read the status display of the compressed air dryer on the chamber rear.
Chambers with voltage and frequency changer: rear wall distance of the alternating climate chamber
approx. 1 m / 3.28 ft to set up the voltage and frequency changer
To completely separate the chamber from the power supply, you must disconnect the power plug. Install
the chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger.
Spacing above the chamber: 100 mm / 3.94 in