Congratulations on purchasing a state-of-the-art Vimmk blood pressure monitor, one of
the most advanced monitors available today. Designed for ease of use and accuracy,
this device will facilitate your daily blood pressure regimen. We recommend that you
read through this manual carefully when the error symbols appear on the screen.
Vimmk
Fully Automatic Upper Arm
Blood Pressure Monitor
Model Number: C04


Replace both batteries with new ones.
Check the battery installation for proper
placement of the battery polarities.
Solution
Weak battery or improper placement
Error Display
No display appears
Cause

Solution
Sensor abnormal
Check if the pump is working or not.
If itis working, then the problem is
sensor abnormal. Please send it to
the local distributor.
Error Display
Err 1
Cause
E r r
1

E r r
2
Error Display
Err 2
Cause
Monitor could not detect pulse wave or
cannot calculate the blood pres sure data.
Solution
Check if the air releasing is too slow or not
If it is too slow, please check if there is any
dust in the tube plug and the cuff port.
If yes, please clean and start the
measuring again. If no, please send the
device back to the local distributor.

E r r
3
Error Display
Err 3
Cause
Measurement result is abnormal.
(SYS≦35mmHg, DIA≦23mmHg)
Solution
Occasionally- repeat measurement.
Always - send it to local distributor.

Error Dis
play
Err 4
Cause
Too loose cuff or air leakage.
(Cannot inflate to 30mmHg within 15s)
Solution
Tie the cuff correctly and make sure the air
plug is properly inserted in the device.
E r r
4

E r r
5
Error Display
Err 5
Cause
The air tube is kinked
Solution
Position the tube correctly and
repeat the measurement.

Solution
The sensor is sensing great
fluctuation in the pressure.
Please keep quiet and don't move.
Cause
E r r
6
Error Display
Err 6

Solution
Cause
The pressure that the sensor
sensing is over the limit.
Please send back to the local distributor.
E r r
7
Error Display
Err 7

Solution
The demarcation is incorrect or the
device has not been demarcated.
Please send back to the local distributor.
Cause
E r r
8
Error Display
Err 8