33
Digital Communication Interface
Interface Commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Secondary Menu
Read proportional band setting pr[op-band] pr pb: 999.9 pb: 15.9
Set proportional band to n pr[op-band]=n pr=8.83 0.1 to 30°C
0.2 to 54°F
Read heater power
(duty cycle)
po[wer] po po: 999.9 po: 6.5
Conguration Menu
Operating Parameters Menu
Read High Limit hl[imit] hl hl: 9999.9 hl: 375.0
Set High Limit hl[imit]=n hl=300 125.0 to 375.0°C
257.0 to 707°F
Serial Interface Menu
Read serial sample setting sa[mple] sa sa: 99999 sa: 1
Set serial sampling setting to n
seconds
sa[mple]=n sa=0 0 to 10,000
Set serial duplex mode: du[plex]=f[ull]/
h[alf]
FULL or HALF
Set serial duplex mode to full
du[plex]=f[ull] du=f
Set serial duplex mode to half
du[plex]=h[alf] du=h
Set serial linefeed mode: lf[eed]=on/of[f] ON or OFF
Set serial linefeed mode to on
lf[eed]=on lf=on
Set serial linefeed mode to off
lf[eed]=of[f] lf=of
Calibration Menu
Read R0 calibration parameter r[0] r r0: 999.999 r0: 100.7
Set R0 calibration parameter to n r[0]=n r=100.7 95.0 to 105.0
Read ALPHA calibration
parameter
al[pha] al al: 9.99999999 al: 0.003865
Set ALPHA calibration parameter
to n
al[pha]=n al=0.003865 0.002 to 0.006
Read DELTA calibration parameter de[lta] de de: 1.50
Set DELTA calibration parameter de[lta]=n de=1.37 de: 9.99999 0–3.0
Miscellaneous Other Commands
Read rmware version number *ver[sion] *ver ver.9999x,9.99 ver.9100S,1.01
Read structure of all commands h[elp] h list of commands
Read ALL operating parameters all all list of parameters
Read setpoint resistance *sr *sr 999.999 ohms 110.023 ohms
Legend: [] Optional command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note: When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed
by a carriage return and linefeed. Then the value is returned as indicated in the RETURNED
column.