9100S Dry-Well Calibrator
Set-point Resistance
22
6.3.1 Scan Control
The scan is controlled with the scan on/off function that appears in the main menu
after the temperature scale units.
Sc Scashesforonesecondandthenthecurrentscansetting
isdisplayed
Off Scanfunctionoff
Press the up or down arrow to toggle the scan on or off.
On Scanfunctionon
Press “SET” to accept the present setting and continue.
S
Acceptscansetting
6.3.2 Scan Rate
The scan rate can be set from 0.1 to 99.9°C/min. The maximum scan rate, however,
is actually limited by the natural heating or cooling rate of the instrument. This rate is
often less than 100°C/min, especially when cooling.
The scan rate function appears in the main menu after the scan control function. The
scan rate units are in degrees Celsius per minute, regardless of the selected units.
Sr Srashesforonesecondandthenthecurrentscanrate
settingisdisplayed
0.1 Scanratein°C/min
Press the up or down arrow to change the scan rate.
2.0 Newscanrate
Press “SET” to accept the new scan rate and continue.
S
Acceptscanrate
6.4 Set-point Resistance
This set-point resistance is used in the calibration calculation of the instrument and
is not adjustable. Once the desired temperature has been reached and the controller
is stable, the set-point resistance can be displayed by pressing “SET” and d
simultaneously. The set-point resistance is displayed as follows.
SrES SrES ashesfortwosecondsandthenthewholenumber
ofthecurrentset-pointresistancesettingisdisplayed
1.888.610.7664 sales@GlobalTestSupply.com
Fluke-Direct.com