94 in t .si g le n t .c o m
Restart The oscilloscope will be triggered when another “Start” occurs before a “Stop".
No Ack The oscilloscope will be triggered when the SDA line is high during any SCL’s ACK bit.
EEPROM
The trigger searches for EEPROM control byte value
1010xxx on the SDA bus. And there is a Read bit and an ACK bit
behind EEPROM. Set the data value and compare type according to
Data1 and Limit Range . If EEPROM’s data is greater than (less
than, equal to) Data1 , the oscilloscope will be triggered at the edge
of the ACK bit behind the Data byte. The Data byte doesn’t need to
follow the EEPROM.
7 Address & Data the oscilloscope will be triggered on the read or write frame in 7-bit address
mode.
Frame (Start: 7-bit address: R/W: Ack) — Data1 and Data2 are set to “0xXX”. If all bits match, then
trigger on the Ack bit followed by the R/W bit.
Frame (Start: 7-bit address: R/W: Ack: Data: Ack: Data2)—If all bits match, then trigger on the Ack bit
followed by the Data2.