MDO-2000EG & 2000EX User Manual
338
Pre-trigger
10 div maximum
Post-trigger
2,000,000 div maximum.
Timebase Accuracy
±50 ppm over any 1 ms time interval
Real Time Sample Rate
1GSa/s max. (4ch models);
1GSa/s per channel (2ch models)
Record Length
Max. 10Mpts
Acquisition Mode
Normal, Average, Peak Detect, Single
Peak Detection
2nS (typical)
Average
selectable from 2 to 256
X-Y Mode
X-Axis Input
Channel 1; Channel 3*
*four channel models only
Y-Axis Input
Channel 2; Channel 4*
*four channel models only
Phase Shift
±3° at 100kHz
Cursors and
Measurement
Cursors
Amplitude, Time, Gating available;
Unit: Seconds(s), Hz (1/s), Phase
(degree), Ratio (%).
Automatic Measurement
38 sets: Pk-Pk, Max, Min, Amplitude,
High, Low, Mean, Cycle Mean, RMS,
Cycle RMS, Area, Cycle Area,
ROVShoot, FOVShoot, RPREShoot,
FPREShoot, Frequency, Period,
RiseTime, FallTime, +Width, -Width,
Duty Cycle, +Pulses, -Pulses, +Edges, -
Edges, %Flicker, Flicker Idx ,FRR, FRF,
FFR, FFF, LRR, LRF, LFR, LFF, Phase.
Cursors measurement
Voltage difference between cursors
(V)Time difference between cursors
(T)
Auto counter
6 digits, range from 2Hz minimum to
the rated bandwidth
Control Panel
Function
Autoset
Single-button, automatic setup of all
channels for vertical, horizontal and
trigger systems, with undo Autoset
Save Setup
20set
Save Waveform
24set
AWG
General
Channels
2
Sample Rate
200MSa/s
Vertical Resolution
14 bits
Max. Frequency
25 MHz
Waveforms
Sine, Square, Pulse, Ramp, DC, Noise
Sinc, Gaussian, Lorentz, Exponential
Rise, Exponential Fall, Haversine,
Cardiac