Insulation Multimeters
Electrical Specifications
37
Model 1587
Output Voltage Display Range Resolution Test Current
Resistance Accuracy
±(% of Rdg + Digits)
0.01 to 6.00 MΩ 0.01 MΩ
50 V
(0 % to + 20 %)
6.0 to 50.0 MΩ 0.1 MΩ
1 mA @ 50 k ±(3 % + 5 counts)
0.01 to 6.00 MΩ 0.01 MΩ
6.0 to 60.0 MΩ 0.1 MΩ
100 V
(0 % to + 20 %)
60 to 100 MΩ 1 MΩ
1 mA @ 100 k ±(3 % + 5 counts)
0.1 to 60.0 MΩ 0.1 MΩ
250 V
(0 % to + 20 %)
60 to 250 MΩ 1 MΩ
1 mA @ 250 k ±(1.5 % + 5 counts)
0.1 to 60.0 MΩ 0.1 MΩ
500 V
(0 % to + 20 %)
60 to 500 MΩ 1 MΩ
1 mA @ 500 k ±(1.5 % + 5 counts)
0.1 to 60.0 MΩ 0.1 MΩ
60 to 600 MΩ 1 MΩ
±(1.5 % + 5 counts)
1000 V
(0 % to + 20 %)
0.6 to 2.0 GΩ 100 MΩ
1 mA @ 1 M
±(10 % + 3 counts)
Model 1577
Output Voltage Display Range Resolution Test Current
Resistance Accuracy
±(% of Rdg + Digits)
0.1 to 60.0 MΩ 0.1 MΩ
500 V
(0 % to + 20 %)
60 to 500 MΩ 1 MΩ
1 mA @ 500 kΩ ±(2.0 % + 5 counts)
0.1 to 60.0 MΩ 0.1 MΩ
1000 V
(0 % to + 20 %)
60 to 600 MΩ 1 MΩ
1 mA @ 1 MΩ ±(2.0 % + 5 counts)