Insulation Multimeters
Cleaning
27
Cleaning
Periodically wipe the case with a damp cloth and mild
detergent. Do not use abrasives or solvents. Dirt or
moisture in the terminals can affect readings.
Testing the Batteries
To test the batteries, press h and turn to the rotary
switch to the INSULATION position. This initiates a battery test
and displays the charge level of the battery.
Testing the Fuse
XWWarning
To avoid electrical shock or injury, remove
the test leads and any input signals before
replacing the fuse.
Test the fuse as described below and shown in Figure 15.
Replace the fuse as shown in Figure 16.
1. Insert a test probe in the
J input terminal.
2. Turn the rotary switch to the
N position and verify
the Meter is in Auto Range.
3. Insert the probe in the mA input terminal. If the
display reading is 0L, the fuse is bad and should be
replaced.
TEST
MIN MAX
HOLD
LO
440 mA
OK
bav14f.eps
Figure 15. Testing the Fuse