FLUKE 1587 Insulation Multimeter

User Manual for the Fluke 1587 Insulation Multimeter - Page 43

For 1587.

PDF File Manual, 48 pages, Read Online | Download pdf file

1587 photo
Loading ...
Loading ...
Loading ...
Insulation Multimeters
Electrical Specifications
33
DC and AC Current Measurement
Range Resolution
Accuracy 1587 and 1587T
±(% of Rdg+Digits)
Accuracy 1577
±(% of Rdg+Digits)
Burden Voltage
(Typical)
400 mA 0.1 mA
±(1.5 % + 2)
[1]
±(2 % + 2)
[1]
AC
45 Hz to
1000 Hz
60 mA 0.01 mA
±(1.5 % + 2)
[1]
±(2 % + 2)
[1]
2 mV/mA
400 mA 0.1 mA
±(0.2 % + 2) ±(1.0 % + 2)
DC
60 mA 0.01 mA
±(0.2 % + 2) ±(1.0 % + 2)
2 mV/mA
[1] 1 kHz bandwidth.
Overload .............................................................600 mA for 2 minutes maximum
Overload Protection.............................................440 mA, 1000 V, FAST fuse
AC Conversion ....................................................Inputs are ac-coupled and calibrated to the rms value of sine wave input. Conversions are
true-rms responding and specified from 5 % to 100 % of range. Input signal crest factor
can be up to 3 up to 300 mA, decreasing linearly to crest factor <= 1.5 at 600 mA. For
non-sinusoidal waveforms add +(2 % reading + 2 % FS) typical, for a crest factor up to 3.