The average number can be set in the selection box after the average sampling mode, and can be set to eight order
of magnitudes: 2, 4, 8, 16, 32, 64, 128 and 256.
The higher the average number is, the slower the response of the displayed waveform-to-waveform changes. A
compromise must be made between the response speed of waveform versus the changes and the degree of noise
reduction shown on the signal.
Use average sampling mode
1) Open the channel menu and select the average sampling mode in the sampling mode option.
2) Press the number box on the right of the average sampling mode box to pop up the average number and then
tap it to set the average number that can best eliminate displayed waveform noises.