Table 16. Troubleshooting (cont.)
Problem Probable Cause Action
Pressure Generation or Indication Problems
Product will
not enter
control mode
Modules not installed
(PCM and 1+
modules)
Verify modules are properly installed.
PMMs not fully
installed
Follow steps outlined in Module Installation to ensure that
all modules are properly installed.
No pressure supply Apply proper supply pressure.
System in remote
operation mode
No pressure
measurement
is shown
PMM not installed Install a PMM (see Module Installation).
Pressure Generation or Indication Problems
Product does
not reach
target
pressure
Pressure supply too
low
Verify pressure supply, perform leak check.
Leak Verify Test port is leak tight.
Valves require
service
Send the Product to Fluke Calibration for repair.
Target pressure set
higher than the
highest ranged PMM.
Install a PMM with the appropriate range.
Slew Rate set to zero Increase slew rate.
While using
measure
mode,
pressure is
leaking
No supply pressure Provide supply pressure if not connected.