Time Measurements
Time measurements include 11 kinds of time parameter measurements.
Figure 81 Time Measurements
1.
Period: Period for every cycle in waveform at the 50% level, and positive
slope.
2. Frequency: Frequency for every cycle in waveform at the 50% level ,and
positive slope
3. + Width: Width measured at 50% level and positive slope.
4. - Width: Width measured at 50% level and negative slope.
5. Rise Time: Duration of rising edge from 10-90%.
6.
Fall Time: Duration of falling edge from 90-10%.
7. BWid: Time from the first rising edge to the last falling edge, or the first
falling edge to the last rising edge at the 50% crossing.
8. + Duty: Ratio of positive width to period.
9. - Duty: Ratio of negative width to period.
10. Delay:Time from the trigger to the first transition at the 50% crossing.
11. T@L: Time from trigger of each transition at a specific level and slope,
include: Current, Max, Min, Mean, and Std-dev.